Gilson 215 Manual do Utilizador Página 54

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 53
3-14
Operation 3
Configuring the Liquid Handler
Safety Contact
The instrument has provisions for connecting safety
devices that your application may require, as long
as they present a contact closure or TTL type
interface. The Safety Contact tab allows you to
specify which input contact is connected to the
safety device and what is the active state of that
device. The function of the safety contact is
equivalent to pressing the Stop button on the
instrument’s front panel.
Bearing Life
The Bearing Life tab displays the XYZ travel in
kilometers.
Vista de página 53
1 2 ... 49 50 51 52 53 54 55 56 57 58 59 ... 111 112

Comentários a estes Manuais

Sem comentários