3-9
Operation 3
Liquid Detector
The Liquid Detector tab allows you to adjust the
default sensitivity for liquid detection.
Note: Gilson control software (such as 735 Sampler
Software) will override these settings.
Raising the percentage suppresses false liquid
detection while lowering the percentage increases
sensitivity for situations where liquid is harder to
detect. By lowering the probe so it touches the
liquid for one of your samples, you can manually
test the current sensitivity of the liquid detection. In
this manner, you can observe the amount of
change you might expect with each sample.
The factory default setting is 10% for each probe.
Configuring the Liquid Handler